Capter 3
Function Setting and Basic Operation
3-8
Curve Tracer CS-3000 Series Instruction Manual
(3) Without patch panel in HIGH VOLTAGE
Remove the patch panel and directly connect a measured device through the attached Test
Lead.
(4) Without patch panel in HIGH CURRENT
Remove the patch panel and directly connect a measure device through the attached Test
Lead.
Be sure to connect to
SENSE terminal for measurement.
If connection to SENSE terminal is not made in HIGH CURRENT mode, no device can be
measured.
Remove the patch panel and directly connect a measured device.
Caution!
♦
Very dangerous voltage is applied to the collector and base terminals. Dangerous operation must be
avoided; i.e., emitter terminal short-circuited with collector or base terminal.
Remove the patch panel and directly connect a measured device.
Test Lead for
HIGH VOLTAGE
(red)
Test Lead for
HIGH CURRENT
(red)
Caution
!
At HIGH VOLTAGE MODE,
connect Collector terminal using
RED color wire which comes with
this unit as standard accessory
and carefully avoid incorrect
wiring.
These connections may cause of
damage on DUT and/or
malfunction of the unit.