Chapter 4 Characteristics Measurement
Curve Tracer CS-3000 Series Instruction Manual
4-13
(7) Press CLEAR key.
(8) Press STEP GENERATOR MENU key.
(9) Press F1 key to select "-STEP."
PNP transistor: minus bias, NPN transistor: plus bias
(10) Press F3 key to select "CURRENT."
(11) Rotate FANCTION knob to set NUMBER OF STEPS to 5.
(12) Rotate STEP AMPLITDE knob to set to
−
200
µ
A/div.
Measurement condition I
C
=
−
100 mA, h
FE
=
70 to 400
in accordance with the performance table.
(13) Press OFFSET knob to set OFFSET to 0.
(13) Press MODE/POLARITY key.
(14) Press F1 key and set to HIGH VOLTAGE.
(15) Press F3 key and set RECTIFIED SINE to "
−
" (
−
full-wave rectified).
(16) Set VERTICAL to 20.0 mA/div and COLLECTOR.
Make proper setting for measurement condition I
C
in accordance with the performance table.
(17) Set HORIZONTAL to 1.00 V/div and COLLECTOR.
Make proper setting for measurement condition V
CE
in accordance with the performance table.