Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-25
2nd menu: How to add the offset in PULSE setting
Note: When MODE / POLARITY is in DC mode, difference by OFFSET continues to be output in STOP.
3.6.4.3
Collector supply functions and related items
Collector supply
MODE
Step generator
STEP
PULSE
AC
Can be set
Can be set
SINE
(full-wave)
Can be set
When set to PULSE, the collector supply
changes to DC.
DC
When set to STEP, the collector supply
changes to SINE (full-wave).
Can be set
LEAK
Can be set for polarity switchover.
Cannot be set
HC
Cannot be set
Can be set
Note 1: When the step generator is changed from STEP to PULSE, the collector supply MODE is changed from
RECTIFIED SINE (full-wave rectified) to DC.
At the same time when the message area on the screen displays "SIN" collector supply selected at
"STEP" at step generator. or "DC" collector supply selected at "PULSE" at step generator. the waveform
icons changes on STEP GENERATOR and COLLECTOR SUPPLY menus.
Note 2: HC: HIGH CURRENT mode is available on CS-3200 and CS-3300.
Submenu
Function description: PREV PAGE - 2nd
F1
OFFSET MODE The offset is added to each pulse.
The offset is added throughout
waveforms.
F6
PREV PAGE
Return to PREV PAGE - 1st menu. Return to PREV PAGE - 1st menu.
Offset
Offset