Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-27
3.7 Setting collector supply
The collector supply generates AC/(SINE) full-wave rectified/DC voltages to be added to the
collector (drain/anode) of a measured device.
3.7.1
VARIABLE knob
3.7.2
MODE/POLARITY key
3.7.3
POWER-WATT key
3.7.4
VOLTS key
3.7.5
ACQUIRE key
3.7.6
LOOPING COMPENSATION key
The setting result is displayed in the setup display area under the waveform display area.
120mW
Max. peak power
30V
Max. peak voltage
2K
Ω
Inner series resistance value
(current limitation resistance)
Mode and polarity
VARIABLE 0.0 % Variable value (output level)
0.0% to 100.0
%
Current limitation resistance
The current limitation resistance value to limit the peak power applied to a measured device is
determined using the maximum peak power and the maximum peak voltage settings as shown
below:
Current limitation resistance