Capter 2
Name and Function of Each Part (Outline)
Curve Tracer CS-3000 Series Instruction Manual
2-13
CAUTION
!
About the caution marks and the voltages that display in the panel and the
patch panel of the test fixture
Missing proper wiring and/or wrong connection may conduct wrong measurement. Also
these connections may cause of damage on DUT and/or malfunction of the unit.
Case1: Wrong connection between Collector terminal and Base terminal;
DUT may be damaged while extra high voltage can be supplied on Base terminal.
Case2: NOT connecting Collector output terminal, Base output terminal and/or Emitter terminal
and connecting Collector sense terminal, Base sense terminal and/or Emitter sense
terminal respectively;
It may cause for burn of SENSE RESISTERs inside of the unit since the sense resisters
connected to outputs respectively for Kelvin sensing and may be supplied current
exceeding each specification.
2.5 Test fixture CS-301
Figure 2.12 shows the patch panel part on the test fixture CS-301 of this instrument and Table 2.11
on next page lists name and function of each part. CS-301 is the test fixture for CS-3100.
Figure 2.12 Name and function of each part on test fixture CS-301