Chapter 4 Characteristics Measurement
Curve Tracer CS-3000 Series Instruction Manual
4-5
4.1.1.2
Measuring diode ON resistance
The diagram on the right in this frame
shows the measurement result of the
diode ON resistance based on the
extracts from the diode performance
table below.
In addition, extracts from the
performance table of device
characteristics which should be
investigated in advance and the
connection diagram indicating how to
connect a measured device to the
attached test adapter are shown.
Diode ON resistance
Caution!
When starting or finishing measurement, operate CS-3000 as shown below:
When starting measurement, set HIGH VOLTAGE breaker to DISABLED and open the test fixture.
After connecting a measured device, set HIGH VOLTAGE breaker to ENABLED.
After finishing measurement, rotate VARIABLE counterclockwise to set the value to 0% (to discharge
applied voltage).
Set HIGH VOLTAGE breaker to DISABLED.
Connection diagram
Connect the anode of the diode to the collector terminal
of the test adapter and the cathode to the emitter
terminal. For connection between the test fixture and
CS-3000, see Section 3.3.2.
Extracts from diode performance table
Absolute max. rating
Peak reverse voltage V
RM
70 V
Peak forward current I
FM
450 mA
Permissible loss
I
O
250mW
Electric characteristics
Item
Measurement Measurement
conditions
value/Max.
Forward
I
F
=10mA
0.8 V
voltage V
F
Test adapter
Patch panel
Anode
(collector)
Cathode
(emitter)