Capter 3
Function Setting and Basic Operation
3-12
Curve Tracer CS-3000 Series Instruction Manual
3.3.7 Connecting the measured device with the test adapter
CS-503/CS-504/CS-505/CS-506 (Option)
[Connection procedure]
(1) Open the cover of the test fixture.
If the cover is opened, this instrument is changed to INTERLOCK state and its output is shut
off.
(2) Connect the test adapter CS-503 (or CS-504, CS-505, CS-506) with the measured device.
1) Insert the terminals of the measured device into the connection grooves of the socket.
CS-503
CS-504
CS-505
CS-506
- For CS-503
:
Mount the measurement device of following package TO-263-3 (D2PAK) on the
socket.
-
For CS-504
:
Mount the measurement device of following package
TO-252-3
on the socket.
-
For CS-505
:
Mount the measurement device of following package
TO-263-7
on the socket.
- For CS-506
:
Mount the measurement device of following package
TO-252-5
on the socket.
The pin assign of the gate terminal and the source terminal corresponds when
the factory is shipped.