(8) Press CONFIGURATION key.
(9) Press F3 key and set to Collector/Short/Common.
(10) Press CLEAR key.
(11) Set VERTICAL to 1.00μA/div and COLLECTOR.
(12) Press ZOOM key and F1 key, and select VERTICAL.
(13) Rotate FUNCTION knob to select "x2".
(14) Set HORIZONTAL to 50V/div and COLLECTOR.
In accordance with the reverse current I
R
and the reverse voltage V
R
of the diode.
(15) Set HIGH VOLTAGE breaker to ENABLED.
(16) Gradually rotate VARIABLE clockwise to apply a voltage to the device.
(17) Press SWEEP key in MEASUMENT part.
For SWEEP measurement, see Section 3.8.3.