Updating software in instrument
Use of SYSTEM TOOLS submenu can update the software programs
*
1
in the instrument.
When updating, attention should be paid to CAUTION below:
∗
1: Normally, our industry expresses the software in this instrument as "Firmware soft" but this instrument
uses "SOFTWARE" in the screen menus and "Software in the instrument" in this document.
The optional software is generically named as "Option soft" in this document and the execution file of
"Option soft" as "Option file".
When updating, follow [Procedure] below:
Caution !
1) When updating the software in the instrument, never set the power supply to standby or remove
the USB memory.
2
)
When the renewal of the software in the instrument is completed and the instrument reactivates, the
renewal of internal hardware might be executed. In this case, please never make the power supply
a standby while updating the software in the instrument.
3) For updating the software in the instrument, contact our sales distributor or our Web site to confirm
adaptability for each device (URL:http://www.iti.iwatsu.co.jp).
4) Updating the software in the instrument causes all the internal settings to be initialized.
Necessary data of SETUP (setting conditions), TRACE (waveform data), and REF (reference
waveform data) should be saved in the USB memory using SAVE function (see Sections 3.11.3
and 3.11.4) in SAVE/RECALL menu. For initialized items, see the default setting items in "Table
3.11.1: Recall function" and "Table 3.11.2 Not recall function" of Section 3.11.8.
5) When saving the file in the software of this instrument in the USB memory, to be sure to enter in
in"FIRMWARE” folder. If not, an error may occur and the software cannot be updated.