2 Installation Manual
- 108 -
(2) RL6 to RL25 (Mini-relay/Single Stable Type)
1) Rating (operation coil)
Class
Rated
voltage
(V)
Rated
Current
(mA)
Coil
resistance
(
Ω
)
Operating
Voltage
(V)
Return
Voltage
(V)
Max.
allowed
voltage(V)
Power
Consumption
(MW)
Single
stable
type
DC12 23.3 514 8.4 1.2
18 Approx.
280
[Table.5]
2) Rating (open-close unit/contact unit)
Class
Single stable type
Number of pole
2
Load
Resistance load
(COS
φ
= 1)
Induction load
(COS
φ
= 0.4, L/R = 7 ms)
Contact type
Crossbar twin
Contact material
Agpd + Au clad
Rated load
AC125V 0.3A
DC 30A 1A
AC125A 0.2A
DC 30A 0.5A
Rated current
3A
Maximum contact voltage
AC250V,DC220A
Maximum contact current
1A
2A
Maximum open-close capacity
(reference value)
AC125VA
DC 60W
AC62.5VA
DC 30W
[Table.6]
3) Performance
Item Description
Number of poles
2
Contact resistance
50m
Ω
or less
Operation time
5 ms or less
Return time
3 ms or less
Electric strength
Between coil contacts :
1 minute at AC 2000 V 50/60 Hz
Between different poles:
1 minute at AC 1000 V 50/60 Hz
Between same poles:
1 minute at AC 1000 V 50/60 Hz
Withstand 10 to 55 Hz double amplitude 5 mm
Vibrations
Malfunction 10 to 55 Hz double amplitude 3.3 mm
Mechanical
AC: More than 100,000,000 times
(at no contact load and open-close frequency of
36,000 imes/hour)
Service life
DC
More than 500,000 times
(at rated load and open-close frequency of 1,800
imes/hour)
Ambient temperature
-40 to +70
°
C (no freezing or condensation)
Ambient humidity
35 to 85% RH
[Table.7]
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......