2 Installation Manual
- 118 -
Procedure/Steps Description/Conditions
2. X-ray exposure conditions
(reference value at actory shipping)
X-ray tube voltage :
100 kV
X-ray tube current :
50 mA
Time :
50 msec
X-ray tube focal point
CXDI chest section
X cm (Set according to the grid used)
3. Tube focal point
CXDI chest section
·Align so that the CXDI chest section and X-ray tube form
a right angle at the center.
·The distance is determined by the setting conditions
during installation.
4. Make the settings for the operation unit according to the
exposure procedure.
5. Perform X-ray exposure, and then load the image onto
the CXDI.
6. Output the loaded image to a high-resolution monitor or
DICOM printer.
Setting of X-ray
exposure conditions
<Positioning>
Positioning between the
X-ray tube and CXDI
Setting of the
exposure conditions
Exposure
Image output
(1)
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......