2 Installation Manual
- 120 -
2) Grid chart check
Procedure
Securely attach the grid chart to the CXDI imaging unit in the same way as the
resolution chart, and then perform steps (1) to (8).
The X-ray exposure conditions are the same as for the resolution chart.
* Blurring and difference in aspect ratio can result if the grid chart is sticks out from the
imaging unit.
Standards
There is no blurring or difference in aspect
ratio in the output image.
(2) Connection with the DICOM printer
1) Overview
The CXDI has been designed for connection to printers from a wide variety
manufacturers, including those already installed and newly added.
Be aware that the final image quality may vary slightly based on the printer performance
and specifications.
Although the following DICOM printers are recommended for connection with the
CXDI, generally, any DICOM-certified printer can be connected to the CXDI.
Also, be sure to inquire at the respective PACS manufacturer when printing with PACS.
In some cases, old non-DICOM printers may be supported.
+. Kodak MLP190
· Grayscale: 4096 x 5120 12-bit printer
+. Imation DryView8700
· Grayscale: 4096 x 5120 12-bit printer
There is no burring or difference in
t ti i th
id h t
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......