4.3.8. Entering Control PC Serial Number
(1) Purpose
To enter the product serial number for “Device Serial Number” of DICOM header.
(2) Procedure
1) Start up the CXDI system.
2) After the exposure screen appears on the operation unit, use the keyboard to enter Debug
mode. ( Use [Alt]+[Tab].)
3) “Welcome to CCR” screen appears. Select “1. Set-Up…”
4) The screen displays “Setting Mode (0: Normal, 1: Expert) [0=0
×
0]:” Select “0: Normal”
Note: Do not select “1: Expert”
5) “CCR SETUP MENU” appears. Select”1. System Setup”
6) “CCR Serial Number [0=0
×
0] : ” appears. Enter the six-digit number indicated on the
naming label of the control PC unit. Press [Enter] key until “CCR SETUP MENU”
appears. [Fig 1]
[Fig 1]
Enter here