3. CXDI-31 Imaging Unit
- 22 -
(4)
Remove the seven screws (XA1-1260-406) and the seven washers (XD1-1102-625) to
remove the CST AD-IF board from the sensor unit.
[Fig.3]
(5)
Attach the new CST AD-IF board to the sensor unit. Temporary tight the seven screws
(XA1-1260-406) and the seven washers (XD1-1102-625) so that the CST AD-IF board
can move.
(6)
At installation of the flat cable (FPC-A-AD), reverse the step (3). Make sure the dust and
any foreign materials are not attached to the connector. Gradually tighten each of the
screws uniformly to prevent the deflection of the connector.
Note:
The flat cable (FPC-A-AD) has multiple pins (120 pins) with high density. Be aware that
the dust in the connector and the deflection of the connector may cause the connection
failure.
Be careful not to touch the TABs
Screws (x7)
Washers (x7)
Gradually tight the screws
uniformly to prevent the deflection
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......