2 Installation Manual
- 83 -
3) Select the END STUDY and transfer the SMPTE pattern image to the printer or
the high-definition monitor. [Fig. 6]
[Fig. 6]
4) On a printed film or on the monitor, measure the density of 32 steps of the grayscale on the
test image. [Fig. 7]
*1: Measure the SMPTE image density on the film using a densitometer. The SMPTE image
density on the high-definition monitor is measured using the gradation analysis software.
*2: The data for the SMPTE test image grayscale transferred by the CXDI are the values for
the maximum density (3.20 in the case of the MLP 190) in the printer settings which have
been changed into 32 uniform steps.
[Fig. 7]
Step
Step32