2 Installation Manual
- 104 -
5. X-ray Controller Interface
5.1. Interface Signal Description
Start X-ray emission
Phototimer
End X-ray emission
X-ray emission signal OFF
Detect X-ray
emission signal ON
Imaging possible?
Enable X-ray
emission signal ON
X-ray generator
CXDI
End image data read
Enable X-ray
emission signal OFF
RX_REQ
RX_COM
X-ray monitor or
setup timer
Generator
setup timer
X-ray emission signal
(1st switch level ON)
X-ray emission signal
(2nd switch level ON)
[Fig.1]
X-ray emission
signal (RX_REQ)
X-ray release signal
(RX_COM)
Actual X-ray emission time
600msec
X-ray generator delay
time (pre-delay)
X-ray generator delay time
(post-delay)
Use edge detection and status detection to
detect X-ray emission signal (XPB)
[Fig.2]
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......