2 Installation Manual
- 99 -
4.3.17. Tool Modes
(1) Purpose
The tool modes (startup options) are intended in order to check operation, and are used to
launch the CXDI application on the control PC by itself, and to display items that are not
normally displayed.
(2) Preparation
1) Connect the keyboard and the mouse to the control PC.
2) Delete the “ccrstart.bat” file from Startup.
3) Disconnect the imaging unit from the control PC.
(3) Notes
Perform the following operation before using the /np mode. Especially, be sure to back up
the exposure mode names and the customized settings before the operation.
1) When using /np with the same settings as that of the connected sensor unit
BodyPart**.ini file can be used as it is.
Example
Connection
/np
setting
Sensor1 Table Table
Sensor2 Stand Stand
2) When using /np with different settings from that of the connected sensor unit
Move the BodyPart**.ini file in the BodyParts folder to the desktop, etc. However, do not
move the Reference folder. If the BodyPart**.ini file is left in the BodyParts folder,
system will not be able to be started, as the sensor type of the BodyPart and the settings
do not match.
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......