2 Installation Manual
- 74 -
Precautions for connecting the server (reference)
Equipment
Restrictions on connections
Kodak
Miil
Transmitting the 0019 shadow group causes a failure, and the group is not
received properly with the default. Its reception is enabled by setting the
strictValidation parameter to Off in Miil.
Fujitsu
Dr. ABLE
y
Transmission is currently performed with the “1 study multi series/1
series 1 image” setting. However, since a multi format is used for the
screen displays for each series under the Dr.ABLE specifications, the
switching operations are a hassle. The user will find it more convenient if
it is at all possible to change the setting to “1 study 1 series/1 series multi
image.” (These unusual data specifications were requested with the full
understanding of their unusualness.)
y
With DICOM, the body parts (such as the abdomen and head) and their
directions (such as PA and AP) belong to the series information. It
therefore follows that a different series is required for a different body
part or body part direction. This aspect is restricted by the DICOM
standard rather than by the installation and other steps taken by us. To
put it the other way around, multiple images with different body parts and
their directions cannot be put together as a series. To remedy this
problem, devising a way of enabling the viewers to reference different
series at the same time at some future point in time will be helpful.
Hitachi
y
Transmission is currently performed with the “1 study multi series/1
series 1 image” setting. However, since a multi format is used for the
screen displays for each series, the switching operations are a hassle. The
user will find it more convenient if it is at all possible to change the
setting to “1 study 1 series/1 series multi image.”
y
This problem arises with the Fujitsu equipment as well. Refer to the
section on Fujitsu.
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......