2 Installation Manual
- 117 -
6. Image Quality Check
(1) Purpose
This procedure is used to check the final image quality of the CXDI.
The settings for the image parameters are made to attain the desired image quality.
1) Resolution check
Tools used
(1) Resolution chart
(2) Grid chart
(3) Image analysis software (currently under development and testing)
(4) High-resolution monitor or DICOM printer
Procedure
Procedure/Steps Description/Conditions
1. Attach the resolution chart to the imaging unit (chest
section) as described below.
Attachment locations: Attach the charts in the
horizontal and vertical directions near the center of
each quadrant
(quadrants 1 to 4).
· The resolution charts are attached securely to the CXDI
chest section.
· Although a single resolution chart is sufficient, the use
of charts in a total of eight locations (one vertical and
horizontal in each quadrant) are recommended.
5.1. Start
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......