2 Installation Manual
- 59 -
(13) -a min_density
-a min_density
Minimum density (Dx100)
Meaning
* This parameter is used in the DICOM Basic Film Box (2010, 0120).
* This parameter designates the density of the digital value for 4095 (0 for
reverse display) of the CXDI transfer data image pixels. (In the CXDI, 4095
indicates white.)
* This parameter is not transferred in many cases since the minimum density
cannot be increased in most printers.
Description
Æ
If this parameter is not designated, it is not transferred over DICOM. In this
case, the printer default values are used. When this parameter is not transferred,
problems can occur since unsuitable default values may be used.
(14) -B border_density
-B border_density
Border density (Dx100) (BLACK/WHITE/D x 100)
Meaning
* This parameter is used in the DICOM Basic Film Box (2010, 0100).
* This parameter determines the area density around the image on the film.
Description
Æ
If this parameter is not designated, it is not transferred over DICOM. In this
case, the printer default values are used.
(15) -G empty_image_density
-G empty_image_density
Empty image density (BLACK/WHITE/D x 100)
Meaning
* This parameter is used in the DICOM Basic Film Box (2010, 0110).
This parameter designates the density of the empty image area during multi-
formatting.
Description
Æ
If this parameter is not designated, it is not transferred over DICOM. In this
case, the printer default values are used.
(16) -T trim
-T trim
Trimming (NO/YES)
Meaning
* This parameter is used in the DICOM Basic Film Box (2010, 0140).
* This parameter adds lines around the image.
* The CXDI is normally adjusted so that the trimming does not appear.
Description
Æ
If this parameter is not designated, it is not transferred over DICOM. In this
case, the printer default values are used.
* For example, the default value for this parameter in Agfa printers is YES.
Therefore, the NO option needs to be specifically designated if it is desired.
(17) -P polarity
-P polarity
Polarity (NORMAL/REVERSE)
Meaning
* This parameter is used in the DICOM Basic Image Box (2020, 0020).
* Reverse image density
Description
Æ
If this parameter is not designated, it is not transferred over DICOM. In this
case, the printer default values are used.
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......