2 Installation Manual
- 91 -
4.3.14. Changing the Total Image Count
(1) Purpose
When the imaging unit is replaced (including the replacement of the LANMIT) for servicing,
the total image count displayed on the user screen can be returned to “0”if necessary.
(2)
Notes
1) The CXDI is connected by the system.
2) Set the CXDI application so that it does not start up.
3) Files are overwritten, and so be careful when performing the procedure.
4) As an extra precaution, write down the numerical values before overwriting them.
5) The overwritten counter becomes valid the next time that the CXDI is started up.
(3) Procedure
1) Turn on the control computer, and then start up Windows NT.
2) Open Windows NT Explorer.
Start
Æ
Programs
Æ
Windows NT
3) A file called “ExpResult.ini” is contained in the CCR folder.
Open this file. (D:\CCR\ExpResult.ini) [Fig 1]
[Fig 1]