Take the SMPTE pattern image again. On the QA screen, compare the image displayed on
the operation unit screen with the film image printed in Step 4) above or with the image on
the high-definition monitor. Make sure there is no difference in contrast and gradation
between those images. Check both preview and magnify images.
If there are any differences between these images, perform the procedure described in “(4)
Operation Unit Image Gamma Correction” on the next page.
If there are no differences between these images, the steps are complete.
(3)
Operation Unit Image Gamma Correction
1) On the QA screen displayed on Step (3) 5) above, select the “Option” tab and press the
“Gamma Adjustment” button. [Fig. 3]
[Fig. 3]