3. CXDI-31 Imaging Unit
- 4 -
(8) LED control
Sensor conditions are shown as [Table.1] below.
Color
Condition
Orange Green
Sleep On
Off
Ready Off
On
Sleep to Ready
Off
Blink
(every 500ms)
Error Off
Irregular blink
(100ms Blink
Æ
100ms Off
Æ
100ms Blink
Æ
500ms Off)
[Table.1]
(9) Serial communication
Performs the serial communication with the Control Station, C2 by using the SCI (Serial
Communication Interface) to control the total operation of the CST AD-IF Board.
(10) Recording logs
Saves the logs related to the total power-on time, number of images taken and the driving
sequence of the LANMIT UNIT to the built-in flash memory.
(11) Remote update
Allows to update the imaging codes and the codes for driving the LANMIT UNIT from the
Control Station C2.
(12) Control line check
Checks whether each control line operates properly according to the commands by the
Capture Board.
2.4. DC/DC Conversion Board
Receives the 14V power from the E/O BOX 31 IF Board and supplies the PCB-AD2
LANMIT UNIT and IC with the driving power.
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......