background image

 

 

 

 

 

 

 

 

Operator’s Manual

 

Summary of Contents for Multi Gas Clip

Page 1: ...Operator s Manual...

Page 2: ...play Layout 5 Display Details 5 Alarms 6 Default Alarms 6 Alarm Behavior 6 Detector Maintenance 7 Bump Testing 7 Manual Bump Test Instructions 7 Calibration 7 Manual Calibration Instructions 7 Battery...

Page 3: ...bustible sensor poisons such as but not limited to alcohol and citrus based cleaners silicones lead compounds eg tetraethyl lead sulfur compounds phosphorus halogenated hydrocarbons and aerosols If su...

Page 4: ...n If the battery is too low for activation the detector will display LOW BAT and then turn off again Once activated the unit will begin to display gas readings immediately Each sensor will show a chas...

Page 5: ...ration bottle will be displayed when it is time to apply gas The battery is displayed in 3 bars as well as a percentage The percentage calculation is approximate and can be used to provide a rough est...

Page 6: ...r s behavior under various alarm conditions Alarm Condition Audible Alarm Vibration Alarm Visual Alarm Low Slow sweep Slow vibration Slow LED Flash High Fast sweep Fast vibration Fast LED Flash TWA Sl...

Page 7: ...on cap provided Once all sensors have gone into alarm the unit will return to normal operation in alarm and the bump due date will be reset If a bump test interval has not been programmed into the uni...

Page 8: ...CT for either warranty replacement or replacement sensor s Battery The battery is displayed as both a percentage and a 3 bar battery icon The percentage calculation is approximate and can be used to p...

Page 9: ...alarm limits Main Menu Date Time Set every time the instrument is communicated via IR Link or Docking Station User Message If the message is empty the detector will skip ahead If the startup message...

Page 10: ...f test lock option specifies whether a button press can silence the alarm Maintenance Notification If maintenance notification is enabled the detector will periodically flash the maintenance LED Other...

Page 11: ...80 days Calibration Gas When the detector is calibrated it scales the sensor readings to match the concentrations of the applied gases The calibration gas concentrations can be adjusted to match the r...

Page 12: ...ese are organized by first in first out FIFO so the 26th event will replace the first event and so on Bump tests are differentiated from normal events when the alarm condition occurs inside of a Clip...

Page 13: ...nty replacement or replacement sensor s BUMP DUE If the detector is displaying BUMP DUE the detector is either due for a bump test because of a scheduled test or has failed its last bump test Refer to...

Page 14: ...ments O2 0 30 vol 0 1 vol increments Sensor Type H2S CO O2 Single plug in electrochemical cell Combustible Plug in Infrared IR or catalytic bead User Options User Message Language Low Alarm High Alarm...

Page 15: ...ble gas detectors you can count on 15 of 16 Contact Information Gas Clip Technologies Inc 218 W Avenue F Midlothian TX 76065 Toll Free 877 525 0808 Phone 972 775 7577 Fax 972 775 2483 E mail info gasc...

Page 16: ...forth in this warranty are conditional on 1 Proper storage installation calibration use maintenance and compliance with the user s manual instructions and any other applicable recommendations of GCT...

Reviews: