2 Installation Manual
- 22 -
4.3.2. LANMIT Image Correction
(1) Purpose
Image data is input to the control PC’s hard disk, and software calibration is performed to
correct the inherent image characteristics of the LANMIT.
(2) Notes
1) This adjustment must be performed during installation, when replacing the imaging unit
(sensor) or the control PC (hard disk), or when the device combination is changed.
2) This adjustment is performed to correct the inherent image characteristics of the
LANMIT. Note that the image data must be entered for each imaging unit that is
connected.
3) This adjustment can be performed from only the control PC. When only the control PC
is started up, a “Sensor Detect Error (-5101)” message is displayed when you open the
CXDI application, but you may ignore this message and proceed with the procedure.
(3) Input of image data file
1) Take off the left-side cover of the control PC.
2) Control the keyboard and mouse to the control PC.
3) Turn on the operation unit, and then turn on the control PC and start up Windows NT.
4) Insert the “Sensor DEF Data” disk supplied with the imaging unit into the floppy disk
drive of the control PC.
5) Open Windows NT Explorer from the Windows NT desktop.
Start
Æ
Programs
Æ
Windows NT Explorer
6) Open”3½Floppy (A:)” and find the “*******.dp” file. Copy this file to the “Ccr” folder
in the D drive.
7) After copying, close Windows NT Explorer, and then remove the “Sensor DEF data”
disk from the floppy disk drive.
8) Exit Windows NT, and turn off the power of all equipment.
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......