2 Installation Manual
- 34 -
5) The sensor information is displayed. Enter the Sensor serial no. and press [Enter].
For CXDI-11/12
serial num =
CH1: BU
MSG:[LIB/DPR/TRN] DPR trn done XXXX
aaaa aaaa
* The serial number entered should be “space + first 4 space + last 4
digits”.
For CXDI-22/31
serial num =
CH1: BU XXXX XXXX XXXX
0400 0001
XXXX XXXX
MSG:[LIB/DPR/TRN] DPR trn done XXXX
aaaa aaaa
* The serial number entered should be “space + first 4 space + last 4
digits”.
6) The screen displays “CH1: RE SN 0000”. Close the CXDI application and return to
Windows NT screen.
7) Restart CXDI application and take the step (4) Sensor serial no. check to ensure the
serial number has been changed.
8) Close HyperTerminal, exit CXDI application and shut down the system.
9) Remove the capture board interface cable.
Sensor serial no. entered
Sensor serial no. entered
Sensor serial no. written is not displayed.
Sensor serial no. entered.
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......