3. CXDI-31 Imaging Unit
- 9 -
Examples of Failure Application
Tape is hang out over the cover edge
Tape does not cover the inner wall of the rim
For the position 2, if you apply the tape in a wrong position, the power lines will be overloaded
by the edge of PCB.
Tape is too long
Tape is too short
Do not use a tool with a thin tip
Power lines will be
overloaded by the edge
of PCB
Wrong position
Correct position of
the tape application