2 Installation Manual
- 112 -
(2) Checking the Pre Delay time stability and setting the internal parameters
* If imaging must be stopped for some reason during the work, select the [Abort] button and
stop the work.
* If the [Modify API] button is pressed, the sequence and parameter data can be changed.
However, this may make accurate measurement impossible, so this button should not be
touched.
A) Start up the [Xpress.exe] file and set the sensor numbers to be used for [Sensor Index] and
the X-ray I/F Channel to be used for [X I/F Ch].
* The Sensor Index and I/F Channel settings should not be changed after selecting the
Phase.
[Screen when starting up the application]
[Fig.3]
B) Select the [Phase1] button to display [Current Script].
[Screen when starting Phase1]
[Fig.4]
Summary of Contents for CXDI-31