2 Installation Manual
- 85 -
Reference 1
In the data and graph below, the data was obtained by setting the Kodak MLP 190 to curve
shape 0 (density linear), a maximum density of 3.20, the SMPTE test image was printed out,
and the image data was measured.
As shown in the graph, the data from the minimum density (film base density) to the maximum
density nearly form a straight line.
SMPTE Density
Step
1
2
3
4
5
6
7
8
9
10
11
Percentage
0
10
20
30
40
50
60
70
80
90
100
Density (D)
0.17
0.46
0.75
1.03 1.36 1.66 1.98 2.29 2.59
2.90
3.20
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
1 2 3 4
5
6
7
8
9
10 11
Step
De
ns
it
y
* The printer parameter settings in this case are:
-A 320 -T NO -M CUBIC -m NORMAL -S CS000