2 Installation Manual
- 116 -
(4) Checking with the measured Pre Delay and Post Delay
When the Phase3 button is selected, the Pre Delay and Post Delay measured in Phase2 are
automatically set.
A) Select the [Phase3] button and select the [Execute] button like Phase2 to set the sensors to
READY status. Then press the [Expose] button on the X-ray generator to irradiate X-rays.
Phase3 ends automatically after 3 irradiations in Phase3.
At this time, check that a value close to the Pre Delay measured in Phase2 is output to
[a], and a value close to the irradiation time set by the X-ray generator is output to [b].
[Phase3 complete screen]
[Fig.9]