2 Installation Manual
- 81 -
6) Create the graph below based on the data measured in step 5).
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
1
2
3
4
5
6
7
8
9
10
11
Step
Density
As shown in the above graph, the measurement values need only to nearly form a straight
line
from the minimum density to the maximum density.
The important point here is that the measurement values make a straight line and a
maximum
density corresponding to the settings is output.
If the measurement values deviate too far from the straight line and a maximum density
corresponding to the settings is not output, printer and monitor output linearity settings,
calibration, and other adjustments are necessary.
* The above graph is an example of measurements when the maximum density is
set to 3.2 D and the printed film is measured with a densitometer. Refer to the
data below when the maximum density is set to 3.1 D, or when gradation analysis
software is used to take measurements on the monitor.
* The LINEAR output cannot be set for some printer models.
In this case, try to select an output setting as close as possible to LINEAR.
Ideal density values in LINEAR LINE
Step
1
2
3
4
5
6
7
8
9
10
11
3.2 D
0
0.32
0.64
0.96
1.28
1.60
1.92
2.24
2.56
2.88
3.20
3.1 D
0
0.31
0.62
0.93
1.24
1.55
1.86
2.17
2.48
2.79
3.10
12 Bit
0
410
819
1229
1638
2048
2457
2867
3276
3686
4095
8 Bit
0
26
51
77
102
128
153
179
204
230
255
Possible causes of non-linearity in measurement values:
Inadequate calibration of printer and high-definition monitor
Inadequate settings for CXDI printer parameters
Faults in printer or high-definition monitor
Max setting
density
Film base
density
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......