2 Installation Manual
- 38 -
4.3.7. Identifying the sensor units and setting the number of units to be connected
(1) Objective
1) In order for the control PC to identify the sensor units connected, the sensor serial number
of each sensor unit is input to the control PC.
2) When the sensor units are connected, the number of the units which have been connected
must be recognized by the control PC, and the settings are performed to this end.
(2) Checkpoints
1) These operations must always be implemented at the installation stage and when any of
the sensor units or control PC (hard disk) has been replaced or when the combination of
equipment has been changed.
2) The sensor serial numbers must always be input even when only one sensor unit is to be
connected. These numbers are the same as what is input to the A/D board. (Refer to
“ Checking the sensor serial numbers”.)
(3) What to have ready
Tool keyboard, tool mouse
(Refer to “System Manual” for the connection methods.)
Setting method
1) Start up the CXDI unit.
2) Once the normal sensor screen has appeared on the operation unit, use the keyboard to
transfer to the debugging mode (by pressing the [ALT] + [TAB] keys).
3) “Welcome to CCR” appears. Select “1 Set-Up...”
4) “Setting Mode (0:Normal, 1:Expert)[0=0x0]:” appears. Select “1:Expert.”
5) “CCR SETUP MENU” appears. Select “7 Scan Sensor Setup.”
6) The “Capture Device Configuration Table” appears. By inputting the number of sensor
units in the underlined section of “Max Capture Devices” below, the control PC is made
to recognize the number of connected sensor units. For example, input “1” when only
one sensor unit is connected, or “4” when four sensor units are connected.
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......