2 Installation Manual
- 110 -
5.4. Adjusting the Timing with the X-ray Generator
(1) Adjusting the Timing
1) Purpose
Adjust the parameters to synchronize the X-ray generator radiation timing with the CXDI.
2) Tools used
· Keyboard
· Mouse
· Timing adjustment software (Xpress)
3) Precautions
· The Xpress software application will automatically be installed if the application is
Version 4.10 or later..
· When connecting two or more X-ray tubess, the timing must be adjusted and the
parameters input for each X-ray tube.
4) Procedures
This work is broadly divided into the following tasks.
A: Preparing for measurement
B: Checking the Pre Delay time stability and setting the internal parameters
C: Measuring with a tentative Pre Delay
D: Checking with the measured Pre Delay and Post Delay
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......