3. CXDI-31 Imaging Unit
- 17 -
3.2.2.2.
Replacing the former sensor cable with the new sensor cable (BG7-3055)
Follow the steps below to replace the old type sensor cable (BG7-2356 or Y67-2676) with the
new type sensor cable (BG7-3055). The cable replacement requires the following parts in set,
which are listed in the Key no. 20, 32, 33, 34 in the [Table.1] below.
[Fig.1]
Parts list
Key No.
Description
Former Part No.
New Part No. Q’ty
Remark
BG7-2356-000
Without CLAMP
20
Cable Unit, Imaging
(Sensor)
Y67-2676-000
With CLAMP
BG7-3055-000
1 With
CLAMP
32 Clamp,
Cable
BA4-0567-020 BA4-2364-000
1
33 Screw
XA1-1260-406 XA1-1260-606
2 M2.6x6mm
34 Washer
-
XD1-1102-625
2 Newly
Added
[Table.1]
[Fig.2]
Old type sensor cable
(BG7-2356 or Y67-2676)
20-3
20-2
20-1
20
32
34
33