2 Installation Manual
- 24 -
12) When a DP file is created, “defPat process success.” and “Make Bitmap image of
defect pattern?[y]:” are displayed. Enter “n”.
13) You will go back to the “CCR IP DEBUG CALIBRATION PROCESS MENU”.
Select “20. Edit Text Info. In Defpix File”.
14) The message below appears. From 0 to 3, select the serial number of the sensor for
which you want to correct the DP file.
A/D Board No. 0-0 Serial No. ********
A/D Board No. 0-1 Serial No. ********
A/D Board No. 0-2 Serial No. ********
A/D Board No. 0-3 Serial No. ********
Select Defect Pixel File ([0-3]: AD No. Other :Specify)[0= 0x0]:
15) The message below appears. Carefully select as follows because proper image
correction is not performed if you mistake the selection.
DP file for CXDI-31 ……… Enter “2”. (Seamless correction not required)
DP file for CXDI-11, 22 ….. Enter “1”. (Seamless correction required)
######## Seamless-Type Part Edit ########
Edit (1) or Clear (2) Info. [1=0x1]:
* Only when you enter “1”, “Select Seamless Type 0:No 3:Type3 100:TypeH[3=0x3];” is
displayed. Enter “2”. (Seamless correction required)
16) The message below appears. Enter “2”.
######## Sensitivity-Correction Part Edit ########
Edit (1) or Clear (2) Info. [1=0x1]:
17) “Do you want to save this command? (N:0, Y:1) [1=0x1]” is displayed. Enter “1”.
18) “save defect pattern files Path [********.dp]:” appears. Enter the DP file to which the
correction data is to be written. Usually the file name is overwritten and you do not have
to enter the name.
19) “save success.” is displayed. You will go back to the “CCR IP DEBUG
CALIBRATION PROCESS MENU”. Move to “Welcome to CCR” menu and select “8.
Exit” to close the CXDI application.
Summary of Contents for CXDI-31
Page 11: ...Labels and Markings Rear iii ...
Page 14: ...Labels and Markings Rear vi ...
Page 18: ......
Page 20: ......
Page 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Page 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Page 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Page 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Page 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Page 184: ......
Page 186: ......
Page 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Page 198: ......