80 - User Manual - RAB108EA
Place the control vial in the appropriate position
of the tube holder and close the TUBE HOLDER
DOOR. (The Cap of the ABX DIFFTROL allows for cap
piercing). If you do not want to pierce the cap, re-
move it and proceed by placing the sample and
closing the tube holder door.
When the tube holder door re-opens, remove (and
recap) the vial.
If any parameter results and/or any statistical
data are displayed in red, perform the following
steps:
1- Verify that the analyzed QC results correlates
with control lot, (scrolling list). If they do not
match, perform a control identification procedure
(See Section 3,
2- Rerun the control material.
3- Perform a concentrated cleaning on the system
(See Section 7,
then rerun the control material again.
Exit the «Quality Assurance» screen