Pentra
80
7–26
Pentra
80 - User Manual - RAB108EA
Fig. 7–31 Super User\Mechanical\Check Sensors
Verify that all sensors are indicated in «Green». If any sensor(s) are indicated in «Red», con-
tact your local
ABX DIAGNOSTICS
Technical Support Representative for further instructions.
5.1.5. Sampler test
Obtain a Sample tube rack and place some tubes in position and leave 2 or 3 empty spaces.
rom the Main screen, select the «Super User Menu» and then
/Mechanical /Sampler Test.
Now select the «Start Rack» key and verify that the rack identification and all the tubes in
the rack have been detected:
◆
Rack number
◆
Type of rack (CBC or DIFF)
◆
Barcode numbers