Maintenance
& Troubleshooting
Troubleshooting
Pentra
80 - User Manual - RAB108EA
7–41
tive replace the valve.
e – If this does not correct the HGB results, call your ABX Diagnostics representative de-
partment.
Differential (LMNE):
a – Check optical bench lamp is lit when instrument is on. If not replace the lamp
2.2. Optical bench lamp replacement,
b – Run a Cytometer rinse (see
c – Re-run the specimen.
d – If this does not correct the WBC results, call your ABX Diagnostics representative de-
partment.