Maintenance
& Troubleshooting
Error messages
Pentra
80 - User Manual - RAB108EA
7–45
8.2. Transfer error types and help messages
Sampling syringe mechanism
not reaching home
Sampling Syringe Motor Mecha-
nism initialization failed
Run an Auto Clean
Check Motor in Service Menu
Thermostated Compartment
Temperature sensor not con-
nected
Thermostated Compartment Tem-
perature
Run an initialization
Thermostated Compartment
Temperature sensor failure
Thermostated Compartment Tem-
perature
Run an initialization
Thermostated Compartment
temperature out of range. Value
Min. & Max.
Thermostated Compartment Tem-
perature
Run an initialization
Thermostated Compartment
Panel open
Thermostatic Compartment Door
Run an Auto Clean
Alarm
Error Type
Help Message
Loading motor mechanism not
reaching home
Sampler Loading Motor Mechanism
initialization failed
Run an initialization
Check motor in Service Menu
Loading motor failure
Sampler Loading Motor home
switch always detected
Run an initialization
Check motor in Service Menu
Stop rack loading switch not
detected
Sampler Loading Motor
Run an initialization
Check motor in Service Menu
Stop rack loading switch
detected
Sampler Loading Motor
Run an initialization
Check motor in Service Menu
Sampler Transfer mechanism
not reaching home
Sampler Transfer Motor Mechanism
initialization failed
Run an initialization
Check motor in Service Menu
Sampler Transfer motor failure
Sampler Transfer Motor home sen-
sor always detected
Run an initialization
Check motor in Service Menu
Stop rack Transfer switch not
detected
Sampler Transfer Motor End trans-
fer rack switch not detected
Run an initialization
Check switch in Service Menu
Stop rack Transfer switch
detected
Sampler Transfer Motor End trans-
fer rack switch always detected
Run an initialization
Check switch in Service Menu
Mixer mechanism not reaching
home
Mixer Motor Mechanism initializa-
tion failed
Run an initialization
Check motor in Service Menu
Mixer motor failure
Mixer Motor home sensor always
detected
Run an initialization
Check motor in Service Menu
Mixer Bad grabber position
Mixer Motor Grabers sensor posi-
tion not detected
Run an initialization
Check motor in Service Menu
Mixer Bad grabber position
Mixer Motor Grabers sensor posi-
tion detected
Run an initialization
Check motor in Service Menu
Rack in wrong side position
Rack in the wrong side in loading
area
Set rack in the right side posi-
tion and restart automatic cycle
No rack
No Rack in loading area
No message
Alarm
Error type
Help Message
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...