Quality
Assurance
Patient Quality Control (XB)
Pentra
80 - User Manual - RAB108EA
3–17
2.3. XB Data Grid screen
From the «XB graphs» menu (see
page 3-16), select the «Details» key to open the
«XB Data Grid» screen (see
Fig. 3–15 XB Data grid screen
The XB data grid contains the individual hematology data as well as the batch number. The
running date and limit values are also included.
The Statistical data includes the mean of all the batches, Standard deviation and Coefficients
of variation. When the Batch values and/or Means are not within their specific limits
(see
page 3-19), they will be displayed in
Red
if they are greater than their
upper limit, in
Blue
if they are lower than their lower limit.
Select the «XB» key to display the contents within a specific batch.
▼
Printing XB Data grid
Select the «Print/Transmit» key to print a list of the batches.
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...