Quality
Assurance
Quality control
Pentra
80 - User Manual - RAB108EA
3–13
1.7. Running control blood
1.7.1. Running in STAT mode
Verify that the lot of control blood has been entered as described in
Prepare the Blood control according to the specific instructions detailed in the blood control
package insert (temperature, mixing...).
Select the «Stat» key: the tube holder door will open.
Read the lot # with the external barcode reader (or manually type the Lot #)
Select the «OK» key
Place the vial into the appropriate possition and close the door to the backwards position.
When the analysis is complete, the QC results will automatically be saved in the QC menu for
that specific lot #.
1.7.2. Running in Rack mode (With barcode label identification)
Verify that the lot of control blood has been entered as described in
Prepare the Blood control according to the specific instructions detailed in the blood control
package insert (temperature, mixing...).
Position the blood control in any rack (verify that the Barcode label will be readable by the
internal barcode reader).
Place the rack in the rack loader and select the «Start Rack» key.
When the analysis is complete, the QC results will automatically be saved in the QC menu for
that specific lot #.
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...