5.3.11. Statistical function flags
▼
XB flag
This alarm is a specific alarm that is associated with patient quality control. This particular
flag is noted when batch results are outside the XB limits established by the user
.
◆
If one of the mean values from 1 batch of 20 samples is outside of the established limits,
an XB alarm will be activated Section 3,
◆
Selecting and Deselecting an analysis can activate the XB alarm.
◆
The user has the option of enabling or disabling the XB alarm in the «Instrument Settings»
functions. Section 5,
◆
The XB alarm can stop the instrument operations if the user has selected the XB alarm con-
ditions in the «Instrument Settings»functions Section 5,
▼
QC failed
This flag will appear when the quality control does not meet the criteria that were selected
in the «Instrument Settings» functions. The user has the option of selecting or de-selecting
these criteria Section 5,
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...