Work
flow
Run results and associated Flags
Pentra
80 - User Manual - RAB108EA
4–39
5.3.6. Flags on RBC histogram
5.3.7. Flags on PLT histograms
Fig. 4–33 RBC histogram
Standard values for MIC: %5
Standard values for MAC: %45
8.3. Alarms & Curve thresholds
MIC
and
MAC
flags are generated
when the percentage of cells count-
ed in the microcytic area (MIC) and
macrocytic area (MAC) compared to
the total number of RBCs are above
the set limits for both MIC and MAC
percentages set up by the user.
RBC1
and
RBC2
thresholds define
the microcytic and macrocytic areas
and are calculated according to the
MCV and the RDW from the RBC
histogram.
Fig. 4–34 PLT histogram
The PLT histogram contains 256
channels between 2fL and 30fL. A
mobile threshold (at 25 fL by de-
fault) moves according to the
microcytic RBC’s that are present in
the platelet analysis area.
The PLT flags are generated under
the following conditions:
%MAC
%MIC
RBC1 RBC2
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...