Pentra
80
3–28
Pentra
80 - User Manual - RAB108EA
Fig. 3–25 Calibration result grid
◆
When calibration results are received on the display, they are saved as calibration
data and not received as sample results.
◆
Calibration results are not transmitted to a host computer, they are kept in the
calibration “Results Grid”.
◆
If a calibration cycle is rejected, it will not be displayed or stored. An error mes-
sage will be displayed on the screen indicating that the calibration sample was
rejected.
To remove results from the statistical calculation, select the check box (see
page 3-28) to remove the mark.
To remove Parameters from the Coefficient calculations, select the check box (see
page 3-28) to remove the mark.
When a Coefficient of variation is displayed in
Red
, it is above its limit and that parameter
will fail calibration.
4.3.1. Calibration passed
There are only 2 passing criteria for calibration!
1- The Coefficients of variation must be within their limits Section 5,
3.4. Coefficients of variation ranges
2- The percent differences between the Target value and the Mean values must be less
than 20%.
Instrument allows you to select an automatic calibration. Press the «Automatic calculation»
key.
1
2
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...