3.3. Number of calibration runs
The number of calibration runs that can be entered by the user is from (1 to 99) runs if re-
quested. The Minimum number of runs for good statistical calculations during calibration is
(
5
).
From the «Quality Assurance» window, select the «Edit» key (see
Now select the titled field «
Minimum required sampling number for automatic calibration
»,
and highlight the number in the field.
Modify the number to the number of runs requested for calibration (Remember, the Minimum
is 5).
Now select the «OK» key to confirm your entry.
3.4. Coefficients of variation ranges
Coefficients of variation (CV) applied on statistical calculations, as calibration Section 3,
within run grids Section 3,
3.1. Accessing the Within Run Data Grid
, page 3-20 are mod-
ifiable within this window.
Fig. 5–9
CV default values
Select the «EDIT» key and highlight the figure you want to modify.
Enter the new value and select the «OK» key.
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...