Quality assurance settings
Pentra
80 - User Manual - RAB108EA
5–11
3. Quality assurance settings
3.1. Accessing the QA settings
From the «Settings» window, select the «Quality Assurance» key
Fig. 5–7
Quality Assurance settings
There are three configurations available in this menu:
◆
XB options (see
◆
Number of calibration runs (see
3.3. Number of calibration runs,
◆
Calibration, QC and XB coefficient values (see
3.4. Coefficients of variation ranges,
3.2. XB options
The «XB» function Section 3,
2. Patient Quality Control (XB)
, page 3-14 is based on a BULL
method and includes a calculation on 3 or on 9 (extended) parameters.
The 3 parameters include: MCV, MCH, MCHC.
The 9 parameters include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT.
This function allows the operator to select the XB mode: 3, 9 parameters or «XB» calculation
disabled (Menu XB inaccessible).
◆
Select the «EDIT» key and check the chosen radio key.
◆
Select the «OK» key to confirm your selection.
Fig. 5–8
XB options
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...