Maintenance
& Troubleshooting
Error messages
Pentra
80 - User Manual - RAB108EA
7–47
Lower Piercing Sensor in Wrong
Position
Sensor state
Run an initialization
Mismatch Between the First and
Second Barcode Tube Read on
Rack %d Pos. %d
Barcode
Run an initialization
No diluent in analyzer reservoir
Incorrect level of diluent into
reservoir
Check diluent level and run a
prime diluent cycle
Printer alarm
Printer problem
Check printer
Printer module closed
Software
No message
QC Failed
QC Failed
Ckeck Quality Control data in QC
screen
Rack Moving Left Sensor Error
Sensor state
Run an initialization
Rack Moving Right Sensor Error
Sensor state
Run an initialization
Rack not identified
No read of Rack Barcode Label
Check barcode label
Reagent level too low for daily
workload
Out of Reagent
Check reagent
Reagent level too low to run a
analasys
Out of Reagent
Check reagent and restart auto-
matic cycle
Reagent level too low to run a
rack
Out of Reagent
Check reagent and restart auto-
matic cycle
Result not stored
Software
Run an autoclean
RS232 alarm
RS232 external problem
Check host connection
Sample ID %d already in
progress
Software
No message
LIS communication module
closed
Software
No message
Tube Detection Sensor in Wrong
Position
Sensor state
Run an initialization
Tube Level Detection Sensor in
Wrong Position
Sensor state
Run an initialization
Two racks with same ID %d in
transfer rail
Barcode
No message
Unable to launch print module
Software
No message
Unable to launch SIL communi-
cation module
Software
No message
Unloader Sensor in Wrong Posi-
tion
Sensor state
Run an initialization
Upper Piercing Sensor in Wrong
Position
Sensor state
Run an initialization
Waste container full
Waste Container Full
Empty waste container and
restart automatic cycle
XB failed
XB failed
Check XB in XB screen
Alarm
Error Type
Help Message
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...