Pentra
80
4–48
Pentra
80 - User Manual - RAB108EA
6.1.3. Result Grid information
Use the slider (see
page 4-47) to display all the items in the Result list as indi-
cated below:
- Sample ID Number,
- Rack# and position of the sample tube in the rack,
- Patient information: Patient ID, First name, Gender, Date of birth
- Test performed
- Analysis date
- The following indicators: «DANEPT»
◆
D: means «Defect analysis» (see
page 4-29) displayed for:
1- Results exceeding parameter linear capacity
2- Reject on one or several parameters
3- Incorrect HGB blank reference
4- CO alarm
◆
A: means «Pathology Alarms» (see
page 4-29) displayed for:
1- Flags involving distribution curves
2- Flags involving LMNE matrix thresholds
3- MIC (PLA), SCL, SCH alarms
◆
N: the result is outside the «Normal values»
◆
E: the result is outside the «Extreme values»
◆
P: the result has been printed out.
◆
T: the result has been transmitted to the host computer.
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...