Quality
Assurance
Calibration
Pentra
80 - User Manual - RAB108EA
3–29
Fig. 3–26 Auto calibration validation
Select the «OK» key to confirm the Autocalibration.
4.3.2. «Forced» calibration
This is an indication that the current calibration analysis is not within the acceptable limits.
1- The Coefficients of variation are greater than their limits.
2- The percent difference between the Target value and the Mean values is greater
that 20%The calibration is called «Forced» and a warning message shall request
confirmation to continue.
When running calibration, a warning message appears on the screen stating a «Forced» cal-
ibration has been detected, another message will appear asking you if you wish to continue
the calibration.
You have the option of continuing or rejecting the calibration at this time.
4.3.3. Coefficients of calibration ranges
Check that the coefficients of calibration are within the following acceptable limits. Or else,
please contact your
ABX DIAGNOSTICS
representative department.
Coefficients of
calibration
Minimum
Standard
Maximum
GB
90
137
200
GR
100
225
300
HT
100
220
300
HB
25
40
55
PLA
130
290
400
IDR
0.1
0.35
1
VMP
0.1
1.0
2
LMNE
0.5
0.86
1.5
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...