Maintenance
& Troubleshooting
Error messages
Pentra
80 - User Manual - RAB108EA
7–49
Incorrect pos.motor loading
(%d) Min : %d Max : %d
Loader motor bad position
Run an Initialization
Incorrect pos.motor mixer (%d)
Min : %d Max : %d
Mixer motor bad position
Run an Initialization
Incorrect pos.motor needle (%d)
Min : %d Max : %d
Needle motor bad position
Run an Initialization
Incorrect pos.motor sampling
(%d) Min : %d Max : %d
Sampling motor bad position
Run an Initialization
Incorrect pos.motor tranfer
(%d) Min : %d Max : %d
Translation motor bad position
Run an Initialization
Internal Barcode Error
Barcode internal connection
problem
Run an Initialization
Internal synchronization failed
System stop due to synchroniza-
tion problem
Run an Autoclean
Motor %d is busy
Run an Initialization
Run a new cycle while analyzer
is busy
Analyzer already in cycle
Run an Autoclean
Valve already activated
Run an Initialization
Alarm
Error Type
Help Message
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...